Fitting the source enclosure to the instrument

Required materials
Chemical-resistant, powder-free gloves
To fit the source enclosure to the instrument
Warning:
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
source enclosure to the source if a corona pin is fitted (the pin tip is
sharp).
1.
Using two hands, fit the source enclosure to the two supporting studs on
the source adaptor housing.
Caution:
NanoLockSpray source enclosure, move the sprayer platform out
of the source enclosure before closing it (see
2.
Close the source enclosure.
3.
Connect the source enclosure cables to the instrument's connectors.
The cables and connectors are color coded; the blue-sleeved cable
Tip:
connects to the blue connector and the yellow-sleeved cable to the yellow
connector.
4.
Slide closed the instrument's source interface door.
The source components can be contaminated with
To avoid puncture wounds, take great care while fitting the
To avoid damaging the sample inlet when fitting a
Removal and refitting of the source enclosure
page
4-5).