1.
Close the source isolation valve (see
2.
Grasp the cone gas nozzle handle, and use it to rotate the sampling cone
assembly 90°, moving the handle from the vertical to horizontal position.
3.
Slide the sampling cone assembly out of the ion block assembly.
Ion block assembly
Do not open the isolation valve at any time when the
Important:
sampling cone assembly has been removed from the ion block assembly.
Cleaning the sampling cone assembly
page
107).
Sampling cone
assembly
Cone gas nozzle
handle
TP03131
April 2, 2015, 715004496 Rev. B
TP03132
129