To remove the existing capillary:
Warning:
hazardous, toxic, or corrosive materials, and to avoid
spreading contamination to uncontaminated surfaces, wear
clean, chemical-resistant, powder-free gloves when working
with the probe and source components.
To avoid burn injuries, take great care while performing this
Warning:
procedure when the probe and source are hot.
1.
Remove the probe from the source (see
2.
Retrieve the combined 2.5-mm Allen wrench and cone extraction tool
from its storage location on the source adapter housing.
3.
Use the combined 2.5-mm Allen wrench and cone extraction tool to
remove the 3 probe end-cover retaining screws.
Replacing the IonSABRE II probe sample capillary
To avoid personal contamination with biologically
page
End-cover retaining screws
April 2, 2015, 715004496 Rev. B
65).
185