6 Maintenance Procedures
Removing the StepWave assembly from the source assembly
Required materials
•
Chemical-resistant, powder-free gloves
•
Seal breaker and locator tool
•
StepWave assembly removal and insertion tool
Recommendation:
the end of the StepWave assembly removal and insertion tool.
To remove the StepWave assembly from the source assembly:
Notice:
removing it from the source assembly, use only these tools:
• Seal breaker and locator tool
• StepWave assembly removal and insertion tool
1.
Position the seal breaker and locator tool so that it engages as follows:
•
Its pin with the hole on the adaptor housing directly under the end
of the StepWave ion guide assembly.
•
Its lip behind the StepWave ion guide assembly's ion guide cap.
154
April 2, 2015, 715004496 Rev. B
When not in use, store the seal breaker and locator tool on
To avoid personal contamination with biologically
Warning:
hazardous, toxic, or corrosive materials, and to avoid
spreading contamination to uncontaminated surfaces, wear
clean, chemical-resistant, powder-free gloves when working
with the probe and source components.
To avoid damaging the StepWave ion guide assembly when