6 Maintenance Procedures

Removing the StepWave assembly from the source assembly

Required materials
Chemical-resistant, powder-free gloves
Seal breaker and locator tool
StepWave assembly removal and insertion tool
Recommendation:
the end of the StepWave assembly removal and insertion tool.
To remove the StepWave assembly from the source assembly:
Notice:
removing it from the source assembly, use only these tools:
• Seal breaker and locator tool
• StepWave assembly removal and insertion tool
1.
Position the seal breaker and locator tool so that it engages as follows:
Its pin with the hole on the adaptor housing directly under the end
of the StepWave ion guide assembly.
Its lip behind the StepWave ion guide assembly's ion guide cap.
154
April 2, 2015, 715004496 Rev. B
When not in use, store the seal breaker and locator tool on
To avoid personal contamination with biologically
Warning:
hazardous, toxic, or corrosive materials, and to avoid
spreading contamination to uncontaminated surfaces, wear
clean, chemical-resistant, powder-free gloves when working
with the probe and source components.