6 Maintenance Procedures
1.
Open the access doors to the IntelliStart Fluidics system.
2.
Using a PEEK nut, Super Flangeless™ ferrule, and stainless steel ring,
connect the blue, 1/16-inch, 300-mm, PEEK tubing from the LockSpray
pump to port 7 on the LockSpray selector valve, where you use a long
finger-tight fitting.
LockSpray selector valve
LockSpray pump
PEEK nut, Super Flangeless ferrule, and stainless steel ring:
PEEK nut
Long finger-tight fitting:
228
April 2, 2015, 715004496 Rev. B
A
B
C
Super Flangeless ferrule
Stainless steel ring
A
B
C
C
B
A
Wa ters
B
C
A