To remove the ionKey source:
1.
Prepare the instrument for working on its source (see
Warning:
with the instrument's source enclosure open; the source can be hot.
2.
Remove the iKey separation device from the docking port (see the
ionKey/MS System Guide, part number 715004028).
3.
Quit the MassLynx software.
4.
Power-off the µSample manager.
5.
Disconnect the PSPI cable.
6.
Using the ¼-inch wrench, loosen and disconnect the fluid-waste line and
fluid-inlet lines from the µSample manager.
7.
Disconnect the optional post-column addition (PCA) line from the
Auxiliary Solvent Manager (ASM) Flow Control Module.
8.
Swing open the ionKey source enclosure unit from the source mounting
on the mass spectrometer.
9.
Disconnect the high voltage cable (white) from the high voltage supply
outlet on the mass spectrometer.
10. Disconnect the reference-probe high voltage cable (green) from the
reference-probe power inlet on the mass spectrometer.
11. Disconnect the options cable (blue) from the options port on the mass
spectrometer.
12. Disconnect the fluid-infusion line from the onboard IntelliStart Fluidics
on the mass spectrometer.
13. Carefully lift off the ionKey source module, and store it safely.
To avoid burn injuries, take great care while working
Removing the ionKey source
page
April 2, 2015, 715004496 Rev. B
99).
89