Plumbing the sample fluidics delivery system for NanoLockSpray operation
To plumb the analyte system:
Warning:
hazardous, toxic, or corrosive materials, and to avoid
spreading contamination to uncontaminated surfaces, wear
clean, chemical-resistant, powder-free gloves when working
with IntelliStart Fluidics tubing and fittings.
You must finger tighten all PEEK fittings.
Requirement:
1.
Open the access door to the IntelliStart Fluidics system.
2.
Using the stainless steel tubing, connect the sample pump to port 7 of
the sample selector valve.
Requirement:
You must tighten the fittings appropriately.
A
B
C
Sample pump
To avoid personal contamination with biologically
The steel tubing requires high-pressure seals at each end.
A
B
C
C
B
A
Wa ters
C
B
A
Sample selector valve