6 Maintenance Procedures
15. Carefully push the PEEK union/UNF coupling assembly and capillary
into the probe assembly so that the locating pin on the UNF coupling is
fully engaged in the locating slot at the head of the probe assembly.
UNF coupling locating pin
16. Fit the nebulizer adjuster knob to the PEEK union/UNF coupling
assembly.
17. Finger-tighten the nebulizer adjuster knob onto the probe assembly.
18. Fit the probe gasket and end-cover to the probe assembly.
19. Use the combined 2.5-mm Allen wrench and cone extraction tool to fit
and tighten the 3 end-cover securing screws.
20. Replace the combined 2.5-mm Allen wrench and cone extraction tool in
its storage location on the source adaptor housing.
Notice:
• To avoid damaging the probe heater's electrical connections, do
• Take great care not to damage the electrical connections,
21. Fit the probe heater (see
22. Fit the probe to the instrument (see
23. In the Instrument Console, click API
desolvation gas flows.
190
April 2, 2015, 715004496 Rev. B
not handle them. Hold the probe heater only by its body.
capillary sleeve, or capillary when fitting the heater over the
capillary sleeve.
step 1
Probe assembly locating slot
through
step 3
on
page
page
57).
, to start the probe and
210).