To install the new capillary:
Warning:
hazardous, toxic, or corrosive materials, and to avoid
spreading contamination to uncontaminated surfaces, wear
clean, chemical-resistant, powder-free gloves when working
with the probe and source components.
To avoid puncture wounds, handle the probe with care.
Warning:
1.
Use the sharp knife or PEEK tubing cutter to cut an approximately
60-cm (24-inches) length of red PEEK tubing.
Requirement:
2.
Insert one end of the red, PEEK tubing in the probe inlet connector, and
screw the connector, finger-tight, into the PEEK union.
Doing so ensures a minimum dead volume when fitting the
Rationale:
capillary.
3.
Fit the UNF coupling to the new capillary.
4.
Use the needle-nose pliers to slide a new liner sleeve and ferrule onto
the capillary.
5.
Insert the capillary in the PEEK union, and ensure that it is fully
seated.
To avoid personal contamination with biologically
Cut the tubing perpendicular to its horizontal axis.
Probe inlet connector
PEEK tubing
Replacing the ESI probe sample capillary
TP02671
April 2, 2015, 715004496 Rev. B