6 Maintenance Procedures
Cleaning the ionKey source and connectors
Required materials
•
Chemical-resistant, powder-free gloves
•
Lint-free tissue
•
Water
•
Isopropanol or methanol
•
Source of clean, pressurized air
Notice:
instrument's case.
During normal operation, the ionKey source does not require cleaning. After
repeated use, however, material can accumulate around the fluid-line
connectors or electronic connectors in the iKey-separation-device docking port.
Cleaning the fluid-line connectors and electronic connectors:
To remove buildup from fluid-line connectors:
Gently swab the fluid-line connectors and outer edges of the docking port with
a lint-free tissue saturated with isopropanol or methanol.
254
April 2, 2015, 715004496 Rev. B
To avoid damage, do not use abrasives or solvents to clean the
Electronic connectors
Fluid-line connectors