Fitting the source enclosure to the instrument

Required materials
Chemical-resistant, powder-free gloves
To fit the source enclosure:
Warning:
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
source enclosure to the source if a corona pin is fitted (the pin tip is
sharp).
1.
Using both hands, fit the source enclosure to the two supporting studs
on the source adaptor housing.
Caution:
colliding with the cone and breaking, always retract the stage
before closing the source enclosure door.
2.
Close the source enclosure.
3.
Connect the probe adjuster and options cables (to the instrument's
connectors.
4.
Slide closed the instrument's source interface door.
The source components can be contaminated with
To avoid puncture wounds, take great care while fitting the
For the NanoFlow source, to prevent the sprayer
Removal and refitting of the source enclosure
4-11