Waters Xevo G2-XS QTof Overview And Maintenance Manual page 231

Hide thumbs Also See for Xevo G2-XS QTof:
Table of Contents

Advertisement

7.
Using a long finger-tight fitting, connect 1/16-inch, 1,000-mm,
natural-color PEEK tubing to port 5 of the LockSpray selector valve, and
thread the tubing into the waste port.
Waste port
The liquid waste system collects waste without requiring a fitted
Tip:
connection. Waste drains through the connector at the base of the
instrument and into a waste bottle, see
8.
If the system lacks an optional flow sensor and grounded union, skip to
step
13.
9.
Using a long finger-tight PEEK nut and 1/32-inch PEEK ferrule, connect
the red, 1/32-in, 200-mm, PEEK tubing from port 4 of the LockSpray
selector valve to the left-hand side of the flow sensor.
Replacing IntelliStart Fluidics tubing (Standard configuration)
A
A
B
B
C
C
A
Wa ters
A
April 2, 2015, 715004496 Rev. B
C
B
B
C
page