Waters Xevo G2-XS QTof Overview And Maintenance Manual page 163

Hide thumbs Also See for Xevo G2-XS QTof:
Table of Contents

Advertisement

4.
Repeat
step 1
through
placing the hook through one of the support rod holes.
5.
Place the vessels containing the first ion guide and second ion guide
PCB assemblies in the ultrasonic bath for 20 minutes.
6.
Carefully pour the cleaning solution from the vessel holding the first ion
guide PCB assembly into the holding container, retaining the ion guide
in the vessel.
You can reuse the cleaning solution for one subsequent cleaning.
Tip:
7.
Fill the vessel with deionized water, to rinse the first ion guide PCB
assembly, and then discard the water.
8.
Refill the vessel with deionized water, to rinse the first ion guide PCB
assembly a second time, and then discard the water.
9.
Repeat
step 6
through
10. Fill both vessels with deionized water, ensuring that each ion guide PCB
assembly is immersed completely.
11. Place the vessels containing the first ion guide and second ion guide
PCB assemblies in the ultrasonic bath for 20 minutes.
Cleaning the StepWave ion guide assembly
step 3
for the second ion guide PCB assembly,
Hook
Second ion guide PCB assembly
step 8
for the second ion guide PCB assembly.
April 2, 2015, 715004496 Rev. B
163

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents