5 Installing and removing the ionKey source
Requirement:
uppermost, as shown in the following figure.
Infill panel
Thumbscrew
Onboard fluidics
panel
10. Using two hands, fit the ionKey source enclosure to the two supporting
studs on the source adaptor housing.
11. Swing the source enclosure to the closed position, ensuring it locks into
place.
12. Quit the MassLynx software.
Notice:
• ensure the µSample manager is powered-off before connecting the
data/power cable;
• ensure that the mass spectrometer is in Standby mode.
13. Power-off the sample manager, and ensure that the mass spectrometer
is in Standby mode.
82
April 2, 2015, 715004496 Rev. B
The thumbscrews must be installed with the screw thread
To avoid damaging the ionKey source or mass spectrometer,