Waters Xevo G2-XS QTof Overview And Maintenance Manual page 251

Hide thumbs Also See for Xevo G2-XS QTof:
Table of Contents

Advertisement

Fluid-line connector plate location:
3.
Trace back the fluid line you want to replace, and then disconnect it
from the port on the ACQUITY M-Class System, or from the diverter
valve on the mass spectrometer.
Disconnect the fluid line, as instructed in the supplied
Important:
documentation.
4.
Cut off and discard the capillary tube fitting at the UPLC-instrument
end of the fluid line.
5.
Return to the ionKey source fluid-line connector plate, and gently pull
off the O-ring from around the fluid-line compression fitting.
6.
Grasp the PEEK fluid connector, and draw the fluid line out and away
from the connector plate.
7.
Dispose of all parts of the fluid-line assembly in accordance with
standard laboratory procedures for contaminated vessels and sharps.
Replacing the fluid lines of the ionKey source
Connector plate
April 2, 2015, 715004496 Rev. B
251