Waters Xevo G2-XS QTof Overview And Maintenance Manual page 51

Hide thumbs Also See for Xevo G2-XS QTof:
Table of Contents

Advertisement

To install the low-volume vials
Warning:
hazardous, toxic, or corrosive materials, and to avoid
spreading contamination to uncontaminated surfaces, wear
clean, chemical-resistant, powder-free gloves when working
with the reservoir bottles.
1.
If a standard reservoir bottle is fitted, remove it.
2.
Screw each low-volume adaptor into the manifold and finger-tighten it.
Warning:
fragile, low-volume glass vials, take care when screwing them in,
and never use force.
3.
Screw each low-volume vial into the adaptor.
4.
For each low-volume vial, ensure that the ends of the solvent delivery
tubes are close to, but do not touch, the bottom of the vial (see
:
To avoid personal contamination with biologically
To avoid laceration injuries caused by the shattering of
Preparing the IntelliStart Fluidics system
Low-volume adaptor
Low-volume vial
Solvent delivery tube
April 2, 2015, 715004496 Rev. B
page
52).
51