Fitting the source enclosure to the instrument

Required material
Chemical-resistant, powder-free gloves
To fit the source enclosure:
Warning:
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
the source enclosure to the source if a corona pin is fitted (the pin
tip is sharp).
1.
Using both hands, fit the source enclosure to the two supporting studs
on the source adaptor housing.
Caution:
colliding with the cone and breaking, always retract the stage
before closing the source enclosure door.
2.
Close the source enclosure.
3.
Connect the probe adjuster and options cables to the instrument's
connectors.
The source components can be contaminated with
To avoid puncture wounds, take great care while fitting
For the NanoFlow source, to prevent the sprayer from
Removing and refitting the source enclosure
4-11