6 Maintenance Procedures

Disassembling the StepWave ion guide assembly

Required materials
Chemical-resistant, powder-free gloves
Combined 2.5-mm Allen wrench and cone extraction tool
O-ring removal kit
To disassemble the StepWave ion guide assembly:
1.
Use the combined 2.5-mm Allen wrench and cone extraction tool to
remove the 2 screws securing the first ion guide assembly to the second
ion guide assembly.
First ion guide assembly
158
April 2, 2015, 715004496 Rev. B
To avoid personal contamination with biologically
Warning:
hazardous, toxic, or corrosive materials, and to avoid
spreading contamination to uncontaminated surfaces, wear
clean, chemical-resistant, powder-free gloves when working
with the probe and source components.
Second ion guide assembly
Securing screw