Required materials
•
Chemical-resistant, powder-free gloves.
•
The Xevo G2-XS QTof Fluidics Tubing and Fitting Kit.
This kit contains components for both the sample and LockSpray
Tip:
system plumbing.
•
For the LockSpray reference probe connection,
–
375 mm of 0.005-inch ID red PEEK
–
A long finger-tight fitting
To plumb the LockSpray system:
Warning:
hazardous, toxic, or corrosive materials, and to avoid
spreading contamination to uncontaminated surfaces, wear
clean, chemical-resistant, powder-free gloves when working
with tubing and fittings.
You must finger tighten all PEEK fittings.
Requirement:
Replacing IntelliStart Fluidics tubing (Standard configuration)
To avoid personal contamination with biologically
April 2, 2015, 715004496 Rev. B
227