4.6.4 Fitting the outer source assembly to the instrument

Required materials:
• Chemical-resistant, powder-free gloves
• 2.5-mm hex wrench (supplied with the instrument)
To fit the outer source assembly to the instrument:
Warning:
To avoid personal contamination with biologically hazardous, toxic, or corrosive
materials, and to avoid spreading contamination to uncontaminated surfaces, wear clean,
chemical-resistant, powder-free gloves when performing this procedure.
1.
Ensure that the outer source is assembled:
• Fit the lenses and the spacers into the lens holder (see
holder).
• Fit the PCB holder assembly to the lens holder (see
the lens
2.
Connect the 10-pin PCB feed-through connector to the PCB connector socket on the outer
source.
Figure 4–8: Connecting the 10-pin PCB feed-through connector
10-pin PCB feed-through connector
16-pin source-housing vacuum feed-through connector
PCB feed-through cable (supplies power to the PCB)
3.
Ensure that the 16-pin source-housing vacuum feed-through connector is connected
correctly.
holder).
April 27, 2020, 715005564 Ver. 01 (previously released as Rev. A)
Fitting the PCB holder assembly to
Page 62
Fitting the lenses into the lens