Installing The Lc Source Enclosure On The Instrument - Waters APGC Operator's Manual Supplement

Atmospheric pressure gas chromatography
Table of Contents

Advertisement

Installing the LC source enclosure on the instrument

Required materials
Chemical-resistant, powder-free gloves
To install the LC source enclosure on the instrument:
1.
Using two hands, fit the source enclosure to the two supporting studs on
the source adaptor housing.
Caution:
NanoLockSpray source enclosure, slide the sprayer platform out of
the source enclosure before closing it.
2.
Swing the source enclosure closed.
3.
Slide open the instrument's source interface door.
4.
Connect the source enclosure cables to the instrument's connectors.
The cables and connectors are color-coded; the blue-sleeved cable
Tip:
connects to the blue connector and the yellow-sleeved cable to the yellow
connector.
5.
Slide the instrument's source interface door closed.
2-24
Setting up the APGC
To avoid personal contamination with biologically
Warning:
hazardous, toxic, or corrosive materials, and to avoid
spreading contamination to uncontaminated surfaces, wear
clean, chemical-resistant, powder-free gloves when handling
source components.