6 Maintenance Procedures
Notice:
precautions:
• Handle the device with care; the component parts are fragile.
• Do not subject the device to pressures that exceed 69,000 kPa
(690 bar, 10,000 psi).
• Do not apply electrospray potential to the emitter for an extended
period when no solvent is flowing.
• Do not drop the device.
• Do not immerse the device in liquid.
• Do not freeze or overheat the device. Keep it within the allowed
temperature ranges during operation and in storage.
• Use the device's sheath to protect it when it is not in use.
• Do not bend or pull the capillary connection tubing where it couples
to the ionKey source.
• Avoid excess voltage, which can erode the emitter over time.
• Decompress the device before you remove it from the source.
• Do not touch the electrospray emitter. Doing so can cause it to bend.
Notice:
remove it from the source. For further details, see the iKey Separation
Device Care and Use Manual (part number 720004897).
See
Appendix A
98
April 2, 2015, 715004496 Rev. B
To avoid damaging the iKey separation device, observe these
For storage, flush the device with 100% acetonitrile before you
for safety advisory information.