6 Maintenance Procedures
Replacing the ionKey reference-probe capillary
Replace the ionKey reference-probe TaperTip emitter or capillary if either is
irreversibly blocked, contaminated, or damaged.
Removing the ionKey reference probe
Required materials
•
Chemical-resistant, powder-free gloves
•
Combined, 2.5-mm Allen wrench and cone extraction tool
•
5-mm Allen wrench
•
Source of oil-free, inert gas (nitrogen or argon) for drying (air-drying
optional)
•
Wash bottle containing HPLC-grade (or better) 1:1 methanol/water
•
Large beaker
Warning:
hazardous or toxic materials, always wear chemical-resistant,
powder-free gloves while performing the following procedure.
Warning:
performed on its source before starting the following procedure.
To remove the ionKey reference probe:
1.
Prepare the instrument for work performed on its source (see
2.
Disconnect the reference probe's cable from the instrument's high
voltage connector.
3.
Disconnect the capillary from the instrument's IntelliStart Fluidics
system.
4.
Remove the ionKey source enclosure from the instrument (see
5.
Retrieve the combined, 2.5-mm Allen wrench and cone extraction tool
from its storage location on the source adaptor housing.
202
April 2, 2015, 715004496 Rev. B
To avoid personal contamination with biologically
To avoid electric shock, prepare the instrument for work
page
99).
page
100).