Fitting the source enclosure to the instrument

Required materials
Chemical-resistant, powder-free gloves
To fit the source enclosure:
Warning:
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
1.
Using both hands, fit the source enclosure to the two supporting studs
on the source adaptor housing.
2.
Close the source enclosure.
3.
Connect the probe adjuster and options cables (to the instrument's
connectors.
4.
Slide closed the instrument's source interface door.
The source components can be contaminated with
Removing and refitting the source enclosure
4-9