Waters Xevo G2-XS QTof Overview And Maintenance Manual page 242

Hide thumbs Also See for Xevo G2-XS QTof:
Table of Contents

Advertisement

6 Maintenance Procedures
Tubing dimensions (sample delivery system):
Valve/Port
Sample/1
Sample/2
Sample/3
Sample/4
Sample/5
Sample/6
Sample/7
Diverter/4
Required materials
Chemical-resistant, powder-free gloves
The Xevo G2-XS Fluidics Tubing and Fitting Kit
This kit contains components for both the sample and NanoSpray
Tip:
system's plumbing.
Needle-nose pliers
242
April 2, 2015, 715004496 Rev. B
Connection
Waste reservoir
Wash reservoir
Diverter valve
Reservoir Bottle A
Reservoir Bottle B
Reservoir Bottle C
Analyte pump
Waste
ID
OD
(inch)
(inch)
0.040
1/16
0.020
1/16
0.005
1/16
0.020
1/16
0.020
1/16
0.020
1/16
0.040
1/16
0.040
1/16
Length
Color
(mm)
Natural
1,000
Orange
1,000
Red
200
Orange
680
Orange
680
Orange
680
n/a
500
Natural
1,000

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents