3.
Using a long finger-tight fitting, connect orange, 1/16-inch, 680-mm,
PEEK tubing from port 1 of the LockSpray selector valve, through
tubing guide A, and into reservoir bottle A.
A
B
C
Reservoir bottle A
As the tubing emerges from the tubing guide, thread a
Requirement:
long finger-tight fitting over it. Push the tubing through the left-hand
hole, to the bottom of the reservoir bottle, and tighten the fitting.
Follow these suggestions if you encounter difficulty threading the
Tips:
tubing through the guides:
Thread the tubing upward from the bottom hole.
Using needle-nose pliers, make a 20° bend, 10 mm from the end of
the tubing. Then rotate the tubing, as necessary, as you thread it
through the guide.
Replacing IntelliStart Fluidics tubing (Standard configuration)
A
B
C
B
A
ate rs
W
B
A
April 2, 2015, 715004496 Rev. B
C
C