9.
Connect port 2 of the diverter valve to the source probe, observing the
relevant procedure:
For the ESI probe, see
For the IonSABRE II probe, see
10. Close the access doors to the IntelliStart Fluidics system.
Important:
close the access doors to the IntelliStart Fluidics system.
Plumbing the sample fluidics delivery system for
NanoLockSpray operation
This section explains how to plumb the NanoLockSpray sample delivery
system for ACQUITY UPLC M-Class applications.
NanoLockSpray sample fluidics delivery tubing schematic:
LockSpray reservoir
A
B
3
C
Tube guides
All tubing is PEEK, except for that comprising the connection between the
pump and sample selector valve, which is stainless steel.

Plumbing the sample fluidics delivery system for NanoLockSpray operation

Ensure that the tubing does not become trapped when you
Wash
2
1
7
6
4
5
Pump line
From wash
A
B
C
page
57.
page
62.
Sample selector
valve
2
1
2
3
7
6
4
5
1
3
4
April 2, 2015, 715004496 Rev. B
To reference
sprayer
Diverter valve
Pump line
241