7.
Use the 7-mm wrench to loosen the locknut.
8.
Unscrew the finger-tight PEEK union from the UNF coupling.
9.
Remove the ferrule from the capillary.
10. Remove the capillary from the UNF coupling.
11. Dispose of the capillary and ferrule in accordance with local
environmental regulations.

Installing the new capillary

Required materials
Chemical-resistant, powder-free gloves
Needle-nose pliers
7-mm wrench
Combined 2.5-mm Allen wrench and cone extraction tool
Red PEEK tubing
LC pump
HPLC-grade (or better) 1:1 acetonitrile/water
Capillary
Sharp knife or PEEK tubing cutter
Safety goggles
Replacing the IonSABRE II probe sample capillary
Ferrule
To avoid contaminating uncontaminated
Warning:
surfaces with biologically hazardous, toxic, or corrosive
materials, dispose of the capillary and ferrule
according to local environmental regulations.
April 2, 2015, 715004496 Rev. B
187