Place a suitable waste container below the mass spectrometer.
To avoid distorting the drip tray or causing the drain cup
Notice:
to leak, restrain the drain cup when attaching or removing the
waste line.
2.
Slide a drain line over the barbed fitting of the drain (located at the
bottom, left-hand side of the mass spectrometer, and on the base of the
bottle tray).
Drain line
3.
Route the waste lines to the waste container. If necessary, shorten the
waste tubes so that their ends are above the surface of the waste solvent.
To prevent leakage of biologically hazardous,
Warning:
toxic, or corrosive materials,
• do not crimp or bend drain lines. A crimp or bend
can impede flow to the waste container.
• empty the waste container before the lower end of
the drain tubes are covered by waste solvent.
Connecting liquid waste lines
Drain cup
Barbed fitting
TP02684