3.
Remove the source enclosure (see
Removing the source enclosure aids access to the ion block
Rationale:
assembly.
4.
Close the source isolation valve (see
5.
Use the combined 2.5-mm Allen wrench and cone extraction tool to
unscrew the 4, captive, ion block assembly securing screws.
Cleaning the ion block assembly
page
100).
page
108).
Ion block assembly
securing screws
TP03130
April 2, 2015, 715004496 Rev. B
139