To remove buildup from electronic connectors:
1.
Open the source door.
2.
From the back, gently direct a stream of air from a clean source across
the electronic connectors.
To prevent contaminating the inside of the source, do not direct air
Note:
into the source.
3.
Close the source door.
To clean the outside surfaces of the ionKey source:
1.
Use a lint-free tissue, dampened with water.
2.
Clean the other system components according to the supplied
documentation.
Cleaning the ionKey source and connectors