6 Maintenance Procedures

Fitting the new IonSABRE II probe heater

Required materials
Chemical-resistant, powder-free gloves
IonSABRE II probe heater
To fit the new IonSABRE II probe heater:
Notice:
crush or distort the probe heater's electrical connections, capillary
sleeve, or capillary when fitting the heater over the capillary
sleeve.
1.
Use the probe adjuster knob to adjust the capillary so that it protrudes
slightly from the end of the capillary sleeve.
2.
Carefully slide the probe heater over the capillary sleeve on the probe
assembly.
Notice:
connections, do not twist the heater when fitting it to the probe
assembly.
3.
Fit the probe heater to the probe assembly, ensuring that the heater is
fully seated on the probe assembly.
4.
Fit the probe to the instrument (see
5.
In the Instrument Console, click API
210
April 2, 2015, 715004496 Rev. B
The probe heater is easily damaged. Take care not to bend,
Probe heater connections
To avoid damaging the probe heater's electrical
Capillary
page
62).
, to start the desolvation gas.
TP02670
Capillary sleeve