Unscrew the capillary PEEK coupler, and remove the capillary from the
union.
10. Where appropriate, remove the protective PEEK sleeve from the
capillary for reuse.
11. Dispose of the capillary and TaperTip emitter, observing local
environmental regulations.
12. Remove any shards of glass or other contaminants from the union as
follows:
a.
Using the wash bottle containing 1:1 methanol/water, rinse the
union over the large beaker.
b.
Blow-dry the union with inert, oil-free gas.

Installing the new TaperTip emitter and capillary

Required materials
Chemical-resistant, powder-free gloves
Combined, 2.5-mm Allen wrench and cone extraction tool
TaperTip emitter
375-mm length of 75-µm PEEK-capillary tubing, sleeved for protection
by 360-µm ID PEEK tubing
F-130 "finger tight" nut
1.5-mm Allen wrench
Replacing the NanoLockSpray reference-probe TaperTip emitter or capillary
To prevent contaminating the environment with
Warning:
biologically hazardous or toxic materials, which can be
present on the fused silica capillary dispose of the capillary
in accordance with local environmental regulations.
April 2, 2015, 715004496 Rev. B