6 Maintenance Procedures
6.
Using the combined, 2.5-mm Allen wrench and cone extraction tool,
unscrew the three fixing screws from the top of the NanoLockSpray
reference probe.
Warning:
The NanoLockSpray reference-probe tip consists of 5 cm of
exposed, fused-silica TaperTip emitter, which is sharp and fragile.
7.
Remove the NanoLockSpray reference probe from the probe adjuster
assembly.
8.
Unscrew the TaperTip-emitter PEEK coupler, and remove the TaperTip
emitter from the union.
TaperTip-emitter
outlet
198
April 2, 2015, 715004496 Rev. B
To avoid puncture wounds, handle the probe with care.
Capillary
PEEK coupler
TaperTip-emitter
PEEK coupler
TaperTip emitter
Union
Fixing screw
NanoLockSpray
reference probe
Capillary
Capillary with
PEEK protective sleeve