Replacing the IntelliStart Fluidics tubing

In the event of a blockage in the tubing connections between the IntelliStart
Fluidics system components, you must replace the tubing. The following
procedures explain how to replace the tubing for the LockSpray and sample
delivery systems. They do not, however, address probe connections, which
vary according to your application. For probe connection instructions, see the
relevant procedure:
For the ESI probe, see
For the IonSABRE II probe, see
Fit the tubing according to one of these plumbing configurations:
• LockSpray standard configuration see
• NanoLockSpray configuration see

Replacing the IntelliStart Fluidics tubing

page
61.
page
62.
page
224.
page
241.
April 2, 2015, 715004496 Rev. B
223