5. NOTICE! Ensure that the new diaphragm unit is securely hooked into
the tube assembly with the bottom two flaps.
Hook the new diaphragm unit with the bottom two flaps into the brass
rings (B) on the tube assembly and fix them again with the screw (A).
6. Plug cables L20 and L21 back into connectors X203 (L20), X302
(L21) and X303 (L21) on board DX61.
7. Plug cables L11 and L12 back into connectors X501 (L11) and X101
(L12) on board DX6.
8. Reattach the "Rear tube assembly" cover [ → 41].
B
9.10.2

What has to be done after replacing the diaphragm unit?

IMPORTANT
Use the most current software version available on the ORTHOPHOS
XG3D CD.
1. Switch the unit on.
2. Perform a software update to the current unit software version [ → 62].
3. Switch the unit off.
Wait approx. 1 minute. Then switch the unit on again.
All modules are updated in accordance with the configuration.
IMPORTANT
If error message E7 61 17 is displayed:
Perform service routine S017.17.
4. Perform a complete unit adjustment or calibration [ → 184].
Only for Germany, Austria, Switzerland and France:
5.
Perform an acceptance check without consulting an expert.
9 Repair
9.10 Diaphragm unit
405