FH laser light localizer (Ceph)

Replacing the ceph laser module in the FH laser light localizer
9.7.2.1
(ceph)
1. Remove the covers "Cephalometer cover" and "Outside secondary
diaphragm" [ → 41].
IMPORTANT
Do not under any circumstances remove or move the secondary
diaphragm!
Otherwise the system will require readjustment.
2. Pull the cable of the Ceph laser module off connector X407 on board
DX91 and detach the cable from the cable ties and holder.
3. Loosen the screw (A) and remove the FH laser module (Ceph).
4. Insert the new laser module (FH) and retighten screw (A).
IMPORTANT
Make sure that the lines in the laser module run vertically.
5. Plug the cable of the FH laser module back into connector X407 on
board DX91.
6. Lay the cable in its original position and secure it there with cable ties.
7. Reattach the cover "Cephalometer cover". [ → 41]
8. Next, adjust the FH laser light localizer (Ceph) [ → 396].