Service Manual ORTHOPHOS XG 3D / Ceph
63 03 510 D3352
D3352.076.03.11.02 09.2016
9.13

Combisensor

9.13.1

Replacing the rotation unit

Rotation of the rotation unit
CAUTION
The unit is rotated via a motor drive.
Never turn the rotation unit manually, as this could damage the gearing.
1. Loosen the screws of cables L13 and L13a and remove the connector
and bushing from the connector plate (C).
2. Loosen the cable ties on the ring and remove the motor cable L20
and the two light barrier cables L21 from connectors X203 (L20) and
X303 (L21) on the ring.
9 Repair
9.13 Combisensor