S011: Test Step 24 - Sirona Orthophos XG 3D Ceph Service Manual

Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS XG 3D / Ceph
63 03 510 D3352
D3352.076.03.11.02 09.2016
8.10.2

S011: Test step 24

Dosimetry with pulsed radiation
A dosimeter for pulsed radiation (e.g. Mult-O-Meter 512L) is required for
dosimetry.
1. Attach the Mult-O-Meter sensor in the middle of the combisensor.
2. Call service routine S011.24 [ → 261].
Selection field 1 displays "85kV/35mAs".
3. Initiate the radiation.
Radiation uses 200 pulses and 85kV/35mAs.
Important: If you let go of the release button before the maximum
radiation time has elapsed, radiation is terminated prematurely
and the exposure is interrupted. The actual radiation time is
displayed.
When you release radiation during the cool-down interval, a
countdown of the remaining waiting time is displayed in the
Easypad title bar (automatic exposure blocking).
4. Read the dose shown on the Mult-O-Meter.
5. Exit the service routine [ → 262].
6. Remove the Multi-O-Meter sensor from the combisensor.
8 Perform service routines via the control panel
8.10 S011: Dosimetry (without ring movement)