8 Perform service routines via the control panel
8.3 Basic operating procedures in the service menu
258
8.3.2

Selecting service routines and test steps

Selecting a service routine

8.3.2.1
The service menu must be selected.
➢ Select the desired service routine via the arrow keys in the selection
field 1 (A) and confirm the selection via the service key (B).
If the selected service routine has several test steps, the first
selectable test step is displayed in selection field 2 (test step 1 in the
example).

Selecting a test step

8.3.2.2
The service menu must be selected.
Selecting a test step in XG 3D
The required service routine must be selected.
➢ Select the required test step in selection field 2 with the arrow keys
(C) and confirm your selection by pressing the Service key (B).
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS XG 3D / Ceph
63 03 510 D3352
D3352.076.03.11.02 09.2016