5Dark Current Exposure (3D) - Sirona Orthophos XG 3D Ceph Service Manual

Table of Contents

Advertisement

10 Maintenance
10.11 Test exposures/Test images
488
Dark current exposure (3D)
10.11.2.5
You can use the
1. Call the
2. In the structure tree, under
current exposure"
The menu for creating the
the action window.
3. Click on the
The exposure window opens in SIDEXIS XG. It indicates the
current status of exposure readiness.
4. Press the release button. Press and hold down the button until the
exposure is complete, the preview image is displayed in the exposure
window, and the acoustic signal that indicates the end of the
exposure (double beep) sounds (if it has been configured).
The exposure is displayed in the exposure window in SIDEXIS
XG.
Service Manual ORTHOPHOS XG 3D / Ceph
"Dark current exposure"
"Test exposures/test images"
"Test exposures 3D"
element.
"Dark current exposure"
"Image acquisition"
button.
Sirona Dental Systems GmbH
to check the 3D sensor.
menu [ → 469].
"Dark
, select the
appears in
63 03 510 D3352
D3352.076.03.11.02 09.2016