10 Maintenance
10.10 Checking the tube data
A
-
60kV/8mA
-
0.5s
Performing measurements
Analyzing measurements
Finalizing the measurement
Check whether the fan is functioning
Check whether the temperature sensor
is supplying plausible values
468
5. Call the Service menu and the Service routine S002.3 (see Service
A
Manual).
6. Use the arrow keys (A) in
60 kV/8 mA.
7. Use the arrow keys (A) in
0.5 s.
+
S2
+
3
➢ Initiate the radiation. Hold the release button pressed until the set
radiation time has expired.
➢ Read the radiation time on the Mult-O-Meter.
The value for the radiation time displayed on the Mult-O-Meter
must correspond to the radiation time of 0.5s selected in the
service routine. The permissible tolerance is ± 10 %.
If the measured radiation time does
tolerance, replace the
If the measured radiation time falls within the permissible
tolerance, finalize the measurement.
1. Exit the service routine.
2. Switch the unit on via switch (A) (see also Operating Instructions).
10.10.4

Checking the fan and temperature sensor

➢ Check the function of the fan using service routine S005.4 [ → 275].
If the fan is defective: Replace the fan [ → 412].
➢ Read the temperature in the single tank with service routine S005.5
[ → 276].
If the displayed temperature reading is not plausible: Replace the
tube assembly. [ → 412]
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS XG 3D / Ceph
selection field 1
to select the kV/mA level
selection field 2
to select the radiation time
not
fall within the permissible
tube assembly
(see Service Manual).
D3352.076.03.11.02 09.2016