Sirona Dental Systems GmbH
Service Manual ORTHOPHOS XG 3D / Ceph
B
Check whether the idling rollers (A) are
OK
63 03 510 D3352
D3352.076.03.11.02 09.2016
10.13
Checking the idling rollers
➢ Manually turn the ring (B) and check it for smooth and easy
movement.
If the ring does not move smoothly and easily:
Remove the housing covers and check the idle rollers (A) for dirt
and foreign particles. Clean and remove foreign particles if
necessary.
10 Maintenance
10.13 Checking the idling rollers
A
491