10 Maintenance
10.14 Checking the grounding straps
Check whether the grounding straps
have complete and firm contact
492
10.14

Checking the grounding straps

Grounding strap in the stand
1. Perform a visual and "hands-on" inspection of the grounding straps to
ensure that they have complete and firm contact at the positions
marked.
If the grounding straps do not have correct contact:
Fasten the grounding straps correctly.
2. Perform a visual inspection of the grounding straps for damage.
If the grounding straps are damaged:
Replace the grounding straps.