7 Adjusting and calibrating the unit
7.2 Adjustment and calibration via the "Unit adjustment/calibration" menu
214

Ceph - Primary diaphragm

7.2.1.4
7.2.1.4.1
Automatic adjustment: Ceph - Primary diaphragm (standard)
NOTICE
Adjustment 1
Risk of damage to unit
It is essential that the needle phantom is removed from the bite block
holder on the device before a ceph exposure is taken; otherwise, the
combisensor could collide with the phantom.
The OP XG needle phantom is
unit.
The OP XG ceph adjustment phantom is in the pan slot on the unit
[ → 200].
The ceph sensor is plugged into the ceph slot on the unit.
The ear plug holders on the ceph arm have been moved completely
apart and swung out of the beam direction (ap).
1. In the structure tree, under
Primary diaphragm"
"Ceph - Primary diaphragm"
The
area.
2. Make SIDEXIS XG ready for exposure. [ → 197]
3. Take an exposure (64 kV/16 mA; 6.1 s) [ → 197].
SIDEXIS XG calculates the adjustment values S1, S2, S3, and
S4 automatically from the exposure and enters them in the text
boxes of the
The exposure is displayed in the exposure window in SIDEXIS
XG.
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS XG 3D / Ceph
not
in the pan bite block holder on the
"2D Adjustment"
element (S010.3).
menu is displayed in the action
"Ceph - Primary diaphragm"
D3352.076.03.11.02 09.2016
"Ceph -
, click on the
menu.
63 03 510 D3352