Bite Block Holder; Replacing The Bite Block Holder - Sirona Orthophos XG 3D Ceph Service Manual

Table of Contents

Advertisement

9 Repair
9.9 Bite block holder
402
9.9

Bite block holder

9.9.1

Replacing the bite block holder

1. Loosen the two screws (A) and remove the bite block holder.
For units with an occlusal bite block only:
2.
Unplug the connector of cable L41 from board DX51.
3. Unscrew board DX51 from the front half of the defective bite block
holder and screw board DX51 into the new bite block holder.
4. Replace the silicone rings (B) (see chapter "Replacing the silicone
rings [ → 403]").
5. Plug the connector of cable L41 into the board of the new bite block
holder.
6. Place the bite block holder into the groove of the silicone rings on the
tube bend and fasten it with the two screws (A).
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS XG 3D / Ceph
B
A
D3352.076.03.11.02 09.2016
B
63 03 510 D3352