Removing Board Dx32 - Sirona Orthophos XG 3D Ceph Service Manual

Table of Contents

Advertisement

9 Repair
9.2 Height adjustment motor (M1_4)/spindle
374
9.2.2

Removing board DX32

1. Unscrew the bracket (C). The bracket (C) is located in front of board
DX32.
2. Move the stand to a height of 1260 (control panel display)
3. DANGER! Potentially lethal shock hazard! Disconnect the unit from
the junction box of the building installation.
4. Remove the cable ties (A) from cable L2.
5. Loosen all the left-hand screws (B) of the protective plates (E).
Sirona Dental Systems GmbH
Service Manual ORTHOPHOS XG 3D / Ceph
63 03 510 D3352
D3352.076.03.11.02 09.2016