Setup of an X-ray component

Since it is addressable via the network, the X-ray component can in
principle be activated for X-ray image acquisition by any of the PCs
connected to the network.
The program SiXABCon is used to manage networkable X-ray
components. It can be executed on any PC in the network on which
SIDEXIS XG has been installed.
IMPORTANT: In order to avoid IP address conflicts, you should never
operate several networkable X-ray components in the network using the
same IP address. Each X-ray component needs to be assigned a unique
IP address.
Factory setting of the TCP/IP address of the unit:
192.168.15.240 (subnet mask: 255.255.255.0)
Checking existing IP addresses
To find out whether an IP address already exists in the network, enter the
"PING" function in the input prompt (DOS window).
all
1. Switch on
components) which are being operated in the network.
2. Invoke the input prompt (DOS window) from a network computer.
3. At the input prompt, enter "ping" followed by the address to be
checked and then press the Enter key.
Example: "ping 192.168.15.13"
If a network device responds, then this address has already been
assigned.
network devices (computers, printers, X-ray
4 General operating procedures
4.4 Setup of an X-ray component
73