S017: Test Step 15 - Sirona Orthophos XG 3D Ceph Service Manual

Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS XG 3D / Ceph
63 03 510 D3352
D3352.076.03.11.02 09.2016
8.14.10

S017: Test step 15

Selection field
1
* Factory setting
Activate/deactivate the acoustic signal for end of exposure
1. Call service routine S017.15.
S017_5 XG 3D
Once the service routine has been selected, the code for the
current setting is displayed in selection field 1.
2. Use the arrow keys (A) to select the required setting in selection field
1 (see table).
Once the required setting has been selected, the Memory key
lights up.
3. Save the setting [ → 262].
4. Exit the service routine [ → 262].
8 Perform service routines via the control panel
8.14 S017: Configuration service
Code
Function
00
Acoustic signal indicating the end of the
exposure is disabled
01
Acoustic signal indicating the end of the
exposure is enabled*
317

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents