Location 10: System Hardware - Sirona Orthophos XG 3D Ceph Service Manual

Table of Contents

Advertisement

Sirona Dental Systems GmbH
Service Manual ORTHOPHOS XG 3D / Ceph
Error code
Description
E7 10 01
EEPROM cannot be written.
Error code
Description
E7 10 02
FPGA of DX1 is not addressable.
FPGA = Field Programmable Gate
Array
Error code
Description
E1 10 03
The flash file system must be
formatted.
Occurs after replacement of board
DX11.
Error code
Description
E1 10 04
Flash file system formatting in
progress.
Error code
Description
E1 10 05
Flash file system is not ready for
operation.
Error code
Description
E1 10 07
General or unknown error during
self-test of 3D image data path.
Error code
Description
E1 10 31
Serial number of DX88 on DX11
incorrect.
63 03 510 D3352
D3352.076.03.11.02 09.2016
5.5.3

Location 10: System hardware

Actions required
E7 10 01: XG 3D
● Acknowledge error and repeat procedure.
If the error occurs repeatedly ...
● Replace board DX11.
Actions required
E7 10 02
● Replace board DX1.
Actions required
E1 10 03
● Acknowledge error and repeat procedure.
The flash file system is formatted and error
code E110 04 is displayed.
Actions required
E1 10 04
● Wait until the error code automatically
disappears (approx. 2 - 3 mins).
Actions required
E1 10 05: XG 3D
● Execute service routine S009.4 and
format flash file system.
The contents of the error memory are thus lost.
If the error occurs repeatedly ...
● Replace board DX11.
Actions required
E1 10 07 XG 3D
This error is a sequential fault.
● Restart the unit:
1. Switch off the unit.
2. Wait 1 minute.
3. Switch unit on.
4. Repeat procedure and observe causal
error messages.
Actions required
E1 10 31 XG 3D
● Run service routine S009.8 to synchronize
system data.
NOTICE! Then a complete unit
adjustment and calibration must be
performed.
5 Messages
5.5 List of error messages
see
S. [ → 435]
see
S. [ → 429]
see
see
see
S. [ → 288]
S. [ → 435]
see
see
S. [ → 290]